Deutsch English Francais Espanol Portugues

Manual Calibration Set
CalSet

 

Manual calibration unit for the
Mercury Ultratracer UT-3000

 

Automatic Calibratotr

 

Application

 

UT-3000 Mercury Ultratracer

The static calibration unit is meant for manually calibrating the
UT-3000 Mercury Ultratracer.

 

Manual Calibration

 

UT-3000 Mercury Ultratracer

The calibration unit is connected to the UT-3000 Ultratracer. With a special syringe a defined volume of mercury containing gas is drawn from the calibrator and injected into the calibration port of the UT-3000.

After entering the volume of the injected gas, the UT-3000 will automatically calculate the mercury mass from the volume injected and the calibration cell temperature.

 

Operating principle

 

CalSet drawing

The Mercury Instruments Manual Calibration Unit consists of a specially designed mercury vessel containing elemental mercury of high purity, surrounded by an aluminium jacket cooled by a thermoelectric cooler.

The temperature of the mercury vessel is exactly measured with a high precision sensor (PT 100). The electronic temperature sensor has to be connected to the UT-3000 ULTRATRACER using the cable delivered with the system.

The temperature of the calibration gas and the concentration of mercury are displayed on the UT-3000 screen before a calibration is performed.

 

Manual Calibration Set
(CalSet):
Technical Specifications

 

Purity of contained Hg:

99,999999%

Accuracy of T-Sensor:

0,01 °C

Power supply:

230 VAC / 50 Hz,
110 VAC / 60 Hz (option)

Power consumption:

30 VA

Dimensions:

35 x 20 x 25 cm (W x H x D)

Weight:

approx. 4,7 kg

ISO 9001

As a leading supplier of high precision analytical equipment, we strive at all times to offer top quality solutions. Our products are manufactured according to the ISO 9001 quality regulations.

CalSet cover

Download brochure:

 

Last updated:
2017-12-06